Laboratory and Process Analyzers
Automatic Analysers → NewLab Range • OilLab Range • Thermofat • OnLine Range
Filterability of Lubricating Oils • Ring and Ball • Foaming Tester • Oxidation Stability • Evaporation Bath • RBOT & TFOUT • RPVOT • Noack • Air Release • Pensky Martens • Leonardo Pensky Martens • Golleo Abel+Pensky • Aniline Point • Penetration • R.E.C.C. • Abel • Cleveland • Tag • Tag - 6901 • Reid Vapour Pressure • Ductilometer • Herschel Emulsifying • Saybolt Viscometer • Distillation
OilLab 6560 - Golleo
Abel+Pensky
Abel
EN 924
EN 13736
IP 170
IP 491
IP 492
ISO 1516
ISO 3679
ISO 13736
Subject
Flash point on petroleum products having a flash point between -18°C and 71°C (kerosene and solvents). Suitable for flash point detection on different substances and waste materials, solvents…
Pensky Martens
ASTM D93 procedures A, B, C
DIN EN 22719
IP 34
ISO 2719
Subject
Flash Point on petroleum products, gas oils, fuel oils, lubrificants, biodiesel. Suitable for flash point detection on different substances, waste materials, solvents…
Measuring Abel Principle
The sample is warmed up according to the methods. When the sample reaches the selected test temperature, the shutter is opened and the ignition system introduces itself automatically. If the flash point is reached, the detection is done by an ionisation detector. If not, the shutter closes again and the sample continues to warm up until the next test temperature.
Measuring Pensky Martens Principle
The sample is heated and stirred at specified rates, using one of three defined procedures (A, B, or C). An ignition source is directed into the test cup at regular intervals with simultaneous interruption of the stirring, until a flash is detected.
Measuring Abel Devices
- Measurement of the Flash Point detected by an ionisation detector
- Testing unit equipped with two ignition systems
- Electrical system or flame exposure device
Measuring Pensky Devices
- Sample temperature: PT100 sensor completely made in stainless steel resistant to corrosion and shock resistant
- Bath temperature: PT100 sensor
Measuring Parameters
- Temperatures: in °C / °F
- Measuring range: 0°C … +420°C
- Resolution: 0.01 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
Pressure sensor
- Built-in sensor with automatic correction of results to a barometric pressure of 101.3 kPa automatically performed by the software at the end of analysis
Detection system
- A single multi-detector combines the ionization detector and the thermal detector
Double ignition system
Test Cup
- The cup is made of brass provided with high temperature resistant handle
- Sample level mark
Heating
- Electrical heater
- Equipped with over temperature cut-out
- Electrical heating with heating rates as per procedures A, B, C
Stirrer
- An electric motor drives the stirring of the product
- Stirring speed as per selected procedures A, B, C
Shutter
- Automatic mechanism opening the shutter conform to the methods
Cooling system
- Built-in forced air fan (Pensky Martens ASTM D93)
- Liquid refrigerant controlled by internal solenoid valve (Abel)
Safety Devices
- Automatic diagnostic in case of breakage of the sample temperature probe and thermal sensors
- Automatic fire detection system
- Overheating protection with auto shut off during the test
- Gas valve for closing the gas supply (max 30 mBar), at the end of the test
Fire Extinguisher
- Automatic fire detection and extinguisher system
Software Features
- All analytical parameters recorded
- Customizable analysis parameters and methods
- Customizable results report
- Printable graphs and results
The software includes:
Analysis Menu
- Automatic execution of the analysis in accordance to the selected procedure (Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference as well as costumized procedures)
- Automatic handling of samples with unknown flash point
- Display in real time of all the analysis parameters and status
- Fields for introduction of operator and product name
- Expected flash point temperature programmable
- Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
- Configuration menu with up to 20 preset samples and expected flash point
- Automatic barometric correction of results
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs
- Selectable value displaying: °C / Volt
Calibration Menu
- Automatic calibration of each temperature probe
- Up to 100 calibration points (standard with 5 and dynamic with up to 100 points)
- Programmable calibration frequency with selectable validity period and notice/lock-down at expiration date
- Last calibration date referred to each single probe displayed and relevant data printable
- Display of calibration diagram
Data Utilities
- Archive viewer for files recall
- All analysis stored in Excel® compatible format
- LIMS compatible
Integrated Touch Screen Panel PC
- TFT/LCD 8.4"
- Resolution 1024 × 768, 16.2 M colours
- 2 × USB ports for connection to an external printer
- 1 × ethernet port for LAN and LIMS network
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable
- 3 conductors flexible cable with schuko plug
Ambient Temperature
Dimensions
- width 37 cm
- depth 52 cm
- height 32 cm
Weight
Spare Parts for Abel
- LAB-650/05-13: heater
- LAB-650/05-16: PT100 bath
- LAB-650/06-11: cooling valve
- LAB-650/06-12: insulated tube for connection to external cryostat
- LAB-650/06-21: gas valve
- LAB-650/07-01: electrical ignitor
- LAB-650/07-03: micro switch
- LAB-650/07-04: handle
- LAB-650/07-05: gas ignitor
- LAB-650/08-12: PT100 product
- LAB-650/08-13: detection / ionisation cable
- LAB-650/09-04: gas reducer
- LAB-650/09-05: calibrated brass crucible
- LAB-650/09-06: calibrated brass crucible complete with movement
- LAB-650/09-07: cover cup movement only
- LAB-650/10-04: PCB fuses, box of 10
- LAB-650/10-05: main electronic board
- LAB-650/11-01: silicon tubing, 1 meter
- LAB-650/11-02: stirrer / flexible
- LAB-650/12-01: voltage transformer for ignitor
- LAB-650/20-01: support PT100 Teflon
Accessories / Spare Parts for Pensky Martens
- LAB-6000/05-23: heater collar
- LAB-6000/06-21: gas valve
- LAB-6000/07-01: electrical ignitor
- LAB-6000/07-03: micro switch
- LAB-6000/07-04: handle
- LAB-6000/07-05: gas ignitor
- LAB-6000/08-12: PT100 product
- LAB-6000/08-13: detection / ionisation cable
- LAB-6000/08-14: PT100 Bath
- LAB-6000/09-04: gas reducer
- LAB-6000/09-05: calibrated brass crucible
- LAB-6000/10-04: PCB fuses, box of 10
- LAB-6000/11-02: stirrer belt
- LAB-6000-118: fire extinguisher system
Calibration Tools
- OilLab 80: calibration decade box – PT100 simulator
- OilLab 91: set of connectors and cables for OilLab 6000
Linetronic Technologies SA
Via Onorio Longhi 2 • 6864 Arzo • Mendrisio • Switzerland
tel. +41 91 630 07 03 • fax +41 91 630 07 19
email: info@lin-tech.ch